Case Series
 
Performance of the Transcatheter LotusTM Valve System in Patients at High Risk of Paravalvular Regurgitation
Mounir Riahi MD1, Jean-Bernard Masson MD1-2, Charbel Naim MD1, Jean-François Gobeil MD1, Nicolas Noiseux MD1-2, Louis-Mathieu Stevens MD1-2, Marco Spaziano MD3, Nicolo Piazza MD3, Giuseppe Martucci MD3, Benoit de Varennes MD3, Kevin Lachapelle MD3, Jeannot Potvin MD*1-2

1Centre Cardiovasculaire du Centre Hospitalier de l’Université de Montréal
2Centre de Recherche du Centre Hospitalier de l’Université de Montréal
3McGill University Health Centre

*Corresponding author: Dr. Jeannot Potvin, Centre Hospitalier de l’Université de Montréal 3840 St-Urbain, Montréal, Québec, H2W1T8, Tel : 514-890-8000; Fax : 514-412-7212; Email :jeannotpotvin@hotmail.com

Submitted: 08-24-2015 Accepted: 10-20-2015 Published: 11-19-2015

Download PDF

_________________________________________________________________________________________________________________________

 

Article

 
Abstract

Background

Transcatheter aortic valve implantation (TAVI) with currently available systems has been associated with the potential of paravalvular aortic regurgitation (PVR) and device malpositioning. The LotusTM is a second generation, fully repositionable and retrievable transcatheter heart valve that features an outer membrane specifically designed to prevent PVR.

Method

Patients were selected among TAVI candidates upon evaluation by the heart team. They showed a higher risk for PVR as determined by high excentricity index or heavy annular and valvular calcification on cardiac computer tomography. TAVI procedures were performed by experienced operators in two different sites.

Results

A total of ten patients underwent TAVI with the Lotus valve system. Mean age was 83.4±4.1 years. Average Logistic Euroscore II and STS mortality risk score were 6.6% and 5.7% respectively. Eight of these patients were at high risk of PVR (eccentricity index > 0.25 and/or cover index < 8%). All procedures were successful. The valve was partially resheathed and repositioned in 5 cases. No PVR was observed in 9 patients; one had a moderate leak immediately after implant, which resolved to trace at day one. Three patients required permanent pacemaker implantation.

Conclusion

In our initial experience, the LotusTM Valve System allowed precise positioning and prevented PVR despite high-risk anatomy.

Keywords: Transcatheter aortic valve implantation (TAVI); LotusTM Valve System; paravalvular aortic regurgitation; aortic
stenosis.

Introduction

Transcatheter aortic valve implantation (TAVI) is now a proven and recognized alternative to surgical aortic valve replacement (SAVR) for high-risk or inoperable patients with severe symptomatic aortic stenosis [1-4]. However, current devices and techniques have been associated with a higher risk of paravalvular regurgitation (PVR) compared to SAVR and more-than-mild PVR has been linked to adverse outcomes in TAVI patients [5,6]. Multiple factors have been identified as potential causes for PVR including annular eccentricity, lack of oversizing, non-symmetric distribution and/or high burden of calcification and malpositioning [7-9]. While better patient selection and procedural awareness may mitigate this risk, system improvements enabling precise valve positioning and better annular sealing may improve outcomes. The LotusTM aVlve y sStem B(oston c Sientific, aNtick, AM, USA) s i au l lfy repositionable and retrievable device designed to facilitate precise positioning while its Adaptative SealTM technology aims to minimise PVR. The REPRISE I and II studies [10,11] showed promising outcomes in an all-comers TAVI population but have not specifically enrolled patients with high-risk features for PVR. The LotusTM system received CE mark in October 2013 [12] and is currently available under the Special Access Program in Canada. We hereby describe its main features and present the results of a series of 10 patients, the majority being at high risk of paravalvular regurgitation.

SYSTEM OVERVIEW

The system has two main components: a bioprosthetic aortic valve (sizes: 23, 25 or 27mm) and a delivery system designed for a transfemoral retrograde approach.

The valve implant comprises three bovine pericardial leaflets supported by a braided nitinol frame. A radiopaque marker is located at the vertical centre of the frame for visual guidance during deployment. The Adaptive SealTM is a polymer membrane circling the bottom half of the device designed to reduce PVR by conforming to the irregular surfaces of the calcified native anatomy (Figure 1).
 
Case Fig 21.1
 
Figure 1. Overview of the different components of the LotusTM valve.
Image copyright: Boston Scientific.

The valve is premounted on the delivery system and is loaded within the catheter after adequate rinsing. A handle with two controls (control knob and release collar) allows for deployment/locking of the device by counterclockwise rotation of the knob and unlocking/recapture by a clockwise movement. The distal portion of the delivery system is curved to facilitate crossing of the aortic arch (Figure 2).

Case Fig 21.2
Figure 2. The LotusTMValve delivery system.

Image copyright: Boston Scientific.

As the device is unsheathed, the frame shortens and radially expands to provide anchoring within the aortic annulus. Once anchored, further unsheathing produces mechanical expansion of the valve frame until it is locked into its final conformation. At any time prior to final release, including when it is completely deployed and locked, full recapture and retrieval of the valve is possible. The LotusTM valve leaflets become functional early in the deployment process, providing hemodynamic stability and enabling controlled positioning with no need for rapid pacing.

METHOD

Patient selection

Procedures were performed at the Centre Hospitalier de l’Université de Montréal (CHUM) and the McGill University Health Center (MUHC). High-risk patients referred for management of severe aortic stenosis underwent clinical evaluation and, when appropriate, transthoracic echocardiography (TTE), multisliced computed tomography and coronary angiography. The multidisciplinary Heart Team discussed all patients and reached an agreement for the best available treatment.

Among transfemoral TAVI candidates, patients were considered for LotusTM implantation over commercially available transcatheter heart valves if they presented one or more of the following criteria: higher risk of PVR (elliptic annulus and/ or eccentric distribution of annular or leaflet calcification) (Figure 3), prohibitive risk of emergent sternotomy in case of malpositioning (restrosternal mammary grafts) or potential instability associated with rapid pacing (severe LV dysfunctionor significant non-revascularizable coronary disease).
 
Case Fig 21.3
Figure 3. Examples of CT scan morphometric evaluation.

(a) Patient #1 and (b) Patient #2. Note the ovaloïd shape of the annulus in both patients and the heavy calcification over the non-coronary cusp in patient #2

Procedure overview

The procedure can either be performed under general anaesthesia with transesophageal echocardiography (TEE) guidance or under conscious sedation with TTE imaging as needed. After confirmation of proper puncture site and femoral artery pre-closure, the LotusTM introducer sheath is inserted in the common femoral artery and advanced into the abdominal aorta. Contralateral arterial and venous accesses are obtained for aortography and temporary pacemaker placement, the later being placed as a back-up only. The LotusTM valve is advanced over a preshaped extrastiff SafariTM wire. Correct orientation of the pre-curved delivery catheter  is ensured by positioning the radiopaque marker of the undeployed valve towards the greater curvature of the arch. The catheter is then advanced across the annulus and positioned so that the ventricular end of the frame is at the annular level or slightly inside the outflow tract (Figure 4a). Unsheathing is initiated and early valve function provides continuous hemodynamic stability for careful and controlled positioning. Typically, the central marker should be maintained 4 to 6mm above the annular plane (middle height of the curved segment of a pigtail catheter sitting in the right or non-coronary cusp can be used as a reliable marker). Aortography and echocardiography are used to confirm valve position, coronary patency and absence of significant PVR. After fluoroscopic confirmation that the locking process has been successfully  completed, the valve is released by sliding the release window and rotating the collar clockwise. The nosecone is then resheathed to its original position and the catheter is retrieved through the introducer sheath. The temporary pacemaker is removed within a few hours if no conduction delay is observed.
 
Case Fig 21.4
Figure 4. Deployment steps.

(a) Valve crossing; (b) The initial position was lower than desired; (c) Adjusted position with the marker slightly above the annular plane; Final release aortogram showing no aortic regurgitation (d).

Definitions and Outcome

Device success and early safety were recorded according to the VARC-2 criteria [13]. Aortic annulus eccentricity index was determined using the previously described formula 1 – (Dmin/ Dmax) [7]. Clinical evaluation was performed at discharge and 30 days. Prosthetic valve hemodynamic profile was documented at TTE during index hospitalization.

RESULTS

A total of ten patients underwent TAVI with the Lotus valve system. Mean age was 83.4±4.1 years. Average Logistic Euroscore II and STS mortality risk score were 6.6 and 5.7 respectively. Five patients were deemed at high SAVR risk due to a retrosternal mammary graft. Mean aortic valve area was 0.75±0.12cm2 and mean transvalvular gradient was 49.4±13.9mmHg by TTE. Baseline characteristics are presented in Table 1. CT-Scan evaluation revealed an elliptic annulus in most patients, with an average eccentricity index of 0.24 (Table 2).

General anaesthesia was used in all but one patient. Direct TAVI without balloon valvuloplasty was performed in all patients. Hemodynamic instability was not observed. VARCII defined device success was achieved in all cases and there were no vascular complication. A permanent pacemaker was implanted in three patients.
 
Table 1. Patient characteristics

Case Table 21.1
GFR: Glomerular filtration rate; AVA: Aortic valve area; LVEF: Left ventricle ejection fraction; CAD: Coronary artery disease.
* Aortic regurgitation ˃mild.

In order to optimize an initially low implant (n=4) or correct moderate PVR (n=1), partial resheathing and redeployment was successfully performed in half of the cases (Figure 4). In one patient, moderate PVR was observed upon partial deployment of the valve. It was repositioned three times in a slightly higher position in attempts to obtain better sealing of the annulus and reduce PVR. Moderate AR persisted despite what was judged the highest implantation possible without impingement in coronary flow (Figures 5a and 5b). TTE performed in this patient one day after implant showed reduction of PVR from moderate to trace (Figure 5c).

Case Fig 21.5
Figure 5. Resolution of a moderate PVL after 24h in patient #2. Post release TEE showing a moderate PVL (a,b). A four chamber TTE at 24h showing trace AR (c).

PVR was not seen by echocardiography or angiography in the other 9 patients. In another patient, transient QRS enlargement
was resolved by redeploying the valve in a higher position.

Case Table 21.2
Table 2. Morphometric characteristics.

*Decision was made to undersize the valve because of a small LVOT area and prevision of a deep implantation site due to a low left main ostium position (10mm). $ Excentricity index= 1-(minimal annular diam/ maximal annular diamter). # Cover index= 100 x ((valve diameter- annulus area derived diameter) / valve diameter)). LVOT: Left ventricular outflow tract.

Conduction disturbances were observed in 5 patients, including 3 that required permanent pacemaker placement (Table 3). There was no difference in annulus or LVOT oversizing in patients with or without new PPM implantation (10.2% vs 10.1% and 7.8% vs 9.4% respectively, p=NS for both). Median length of stay was 4.5 days (IQ 3-5) but only 3 (IQ 3-3) in patients who did not require permanent pacing. Follow-up at 30 days was completed in the 10 patients and VARC-2 early safety criteria were met in all (Table 4).

Table 3. Conduction outcomes

Case Table 21.3
LBBB: Left Bundle Branch Bloc; AVB: Atrioventricular Bloc;
PPM: Permanent Pacemaker. AF: Atrial Fibrillation

Table 4. Procedural outcome and early safety
Case Table 21.4
 
¥ According to VARC-2 criteria. * TTE at 24h. & Aortic regurgitation
˃mild.

AVB: Atrioventricular block. PPM: Permanent Pacemaker. TIA: transient ischemic stroke.

Discussion

Our initial experience with the LotusTM valve illustrates potential advantages of this second generation system. As TAVI continues to show promising results in high-risk populations, wider application will require improvements in some of the weaknesses of the current technology. First, PVR prevention is increasingly recognized as a determinant factor with multiple reports suggesting that PVR is correlated with increased late mortality [5,6,14]. Imaging studies have suggested that native valve burden of calcification, an annulus eccentricity index ˃0.25 and a cover index ˂8% are predictors of post-procedural PVR [7-9,15-17]. The Lotus valve showed high rates of circularity 12 months post-implantation in a small CT scan study [18] which, in combination with the Adaptive SealTM, may have contributed to the low rate (1.1%) of significant PVR seen in the Reprise II trial [12]. In our cohort, 8 of 10 patients were at high risk of PVR based on either an eccentricity index over 0.25 (5 patients) and/or a cover index under 8% (8 patients). Despite these high-risk characteristics, none of them had moderate or severe PVR at discharge, including both patients with extreme eccentricity (indexes of 0.35 and 0.29 respectively). This may reflect the predominant role played by the Adaptive SealTM in PVR reduction. The only moderate leak observed during the procedure spontaneously regressed to trace at 24 hour. This may be due to continued dynamic conformation of the Adaptive SealTM to the heavily calcified apposition area. Nevertheless, the absence of significant PVR in our small but high-risk cohort suggest that an integrated PVR prevention system may translate into improved clinical outcome. This will require further investigation in a larger population at high risk of PVR.

Second, precise positioning is crucial to avoid serious complications such as device embolization, coronary obstruction or valve-in-valve (TV-in-TV) deployment. TV-in-TV with first generation systems has been associated with increased 1-year mortality [19]. With the Lotus system, partial resheathing can be performed as many times as needed to achieve the optimal desired position within the native annulus. Prior to final release, full recapture of the valve is possible, allowing for device retrieval if a native leaflet obstructs a coronary ostium or if missizing has occurred. These features contribute to increased procedural control and precision and could explain the absence of ectopic or valve-in-valve deployment in the Reprise trials [10-12]. In our series, the repositioning capability proved useful to optimize implant location in half of the cases, including cases where (a) PVR due to a heavily calcified leaflet was reduced and (b) QRS enlargement was corrected with  repositioning. Although minimizing native valve manipulation is generally desirable, recent literature does not show signals of increased stroke risk with repositionable systems [10-12,20] and the need for repositioning may decrease as operators become more familiar with the valve behavior during deployment.

The current LotusTM system version has potential limitations. First, the delivery sheath profile is higher than other second generation systems, requiring minimal access vessel diameter of 6 and 6.5mm, and alternative access experience is limited to the direct aortic approach thus far. Second, post-procedural pacemaker implantation rate was relatively high in the Reprise program, reaching up to 29.7% [12], similar to what was observed in our cohort. This is likely due to an interaction between the Adaptive SealTM and the conduction system in the LVOT, with oversizing of both annulus and LVOT playing a potential role [21]. Our study could not corroborate this finding as no apparent trend between LVOT or annular oversizing and PPM implantation was observed. This may be due to our small sample size, but alternative hypotheses such as implantation depth or technique should be explored. Further studies evaluating the mechanical interaction with the underlying conduction system and its causative effects of dysrhythmia are warranted to better understand this issue.

In the modern TAVI era, the paradigm is shifting from utilization of one universal system to a more tailored therapy for specific populations. Owing to its PVR-reducing capabilities, hemodynamic stability during deployment and repositionability, the Lotus system adds to the overall TAVI arsenal. The results of the real-world RESPOND registry and the REPRISE III randomized trial comparing the Lotus and CoreValve (Medtronic) systems will provide important insights into the role of this technology in the growing field of TAVI.

Conclusion

The LotusTM Valve System provides a combination of advanced features that allow precise positioning. In our selected population at increased risk, paravalvular regurgitation was consistently prevented. Introduction of next generation technology into established TAVI programs may promote individualized therapy and improve outcomes.

Aknowledgement

We wish to thank Mr Jean Gélinas, nurse coordinator for the CHUM TAVI program for his contribution and support.

Funding source

This study did not receive any internal or external funding.

Disclosures

Dr Jeannot potvin is consultant and proctor for Boston Scientific. Dr Jean-Bernard Masson is speaker for Boston Scientific. Dr Nicolo Piazza is consultant and proctor for Medtronic. Dr. Giuseppe Martucci is proctor for Medtronic. Other co-authors have no potential conflict of interest.

Summary

The LotusTM Valve is a second generation TAVI system specifically designed to prevent ParaValvular Regurgitation (PVR). We describe ten of the first Canadian Lotus patients. They were selected based on higher risk anatomic features for PVR. All patients underwent a successful procedure. Three patients needed pacemaker implantation. At 30 days follow-up no PVR was observed. The LotusTM Valve System allowed precise positioning and prevented PVR despite high risk anatomy.

References

 References

1.Collins PW, Chalmers E, Hart D, Ian Jennings, Ri Liesner et al. Diagnosis and management of acquired coagulation inhibitors: a guideline from UKHCDO. Br J Haematol. 2013, 162(6): 758-773.

2.Knoebl P, Marco P, Baudo F, Collins P, Huth-Kühne A et al. Demographic and clinical data in acquired hemophilia A: results from the European Acquired Haemophilia Registry (EACH2). J Thromb Haemost. 2012, 10(4): 622-631.

3.Rahman M, Shad F, Smith MC. Acute kidney injury: a guide to diagnosis and management. Am Fam Physician. 2012, 86 (7): 631-639.

4.Miyamae K, Otsuka T, Otsuka Y, Nagayoshi M, Hamada Y.Clinical study of bladder tamponade resulting from clots of blood [in Japanese]. Nihon Hinyokika Gakkai Zasshi. 2006, 97(5): 743-747.

5.Collins PW, Hirsch S, Baglin TP, Dolan G, Hanley J et al. Acquired hemophilia A in the United Kingdom: A 2- year national surveillance study by the United Kingdom Haemophilia Centre Doctors’ Organisation. Blood. 2007, 109(5): 1870-1877.

6.Baudo F, Collins P, Huth-Kühne A, Lévesque H, Marco P et al. Management of bleeding in acquired hemophilia A: results from the European Acquired Haemophilia (EACH2) Registry. Blood. 2012, 120(1): 39-46.

7.Sumner MJ, Geldziler BD, Pedersen M, Seremetis S. Treatment of acquired haemophilia with recombinant activated FVII: a critical appraisal. Haemophilia. 2007, 13(5): 451- 461.

8.Neufeld EJ, Négrier C, Arkhammar P, Benchikh el Fegoun S, Simonsen MD et al. Safety update on the use of recombinant. activated factor VII in approved indications. Blood Rev. 2015, 19(Suppl 1): S34-41

9.Lauer S, Westphal M, Fischer LG, Andreas Meißner, Hugo Van Aken et al. Recombinant factor VIIa and factor VIII treatment for acquired factor VIII deficiency: a case of repeated thrombotic endotracheal occlusion in a mechanically ventilated patient. Crit Care. 2011, 15(2): 407.

10.Mannucci PM. Hemostatic drugs. N Engl J Med. 1998, 339(4): 245-253.

11.Kato T, Masui K, Yoshida T, Soma T, Mishina M et al. Acquired hemophilia A developing at bilateral renal bleeding: a case report [in Japanese]. Hinyokika Kiyo. 2009, 55(4): 215-218.

12.Takimoto M, Nagura S, Sano F, Nagoshi H, Miura I. Acquired hemophilia complicated with hemorrhage induced acute renal failure [in Japanese]. Rinsho Ketsueki. 2010, 51(6): 407-412.

13.Rahman M, Shad F, Smith MC. Acute kidney injury: a guide to diagnosis and management. Am Fam Physician. 2012, 86 (7): 631-639.

14.Miyamae K, Otsuka T, Otsuka Y, Nagayoshi M, Hamada Y.Clinical study of bladder tamponade resulting from clots of blood [in Japanese]. Nihon Hinyokika Gakkai Zasshi. 2006, 97(5): 743-747.

15.Collins PW, Hirsch S, Baglin TP, Dolan G, Hanley J et al. Acquired hemophilia A in the United Kingdom: A 2- year national surveillance study by the United Kingdom Haemophilia Centre Doctors’ Organisation. Blood. 2007, 109(5): 1870-1877.

16.Baudo F, Collins P, Huth-Kühne A, Lévesque H, Marco P et al. Management of bleeding in acquired hemophilia A: results from the European Acquired Haemophilia (EACH2) Registry. Blood. 2012, 120(1): 39-46.

17.Sumner MJ, Geldziler BD, Pedersen M, Seremetis S. Treatment of acquired haemophilia with recombinant activated FVII: a critical appraisal. Haemophilia. 2007, 13(5): 451- 461.

18.Neufeld EJ, Négrier C, Arkhammar P, Benchikh el Fegoun S, Simonsen MD et al. Safety update on the use of recombinant. activated factor VII in approved indications. Blood Rev. 2015, 19(Suppl 1): S34-41

19.Lauer S, Westphal M, Fischer LG, Andreas Meißner, Hugo Van Aken et al. Recombinant factor VIIa and factor VIII treatment for acquired factor VIII deficiency: a case of repeated thrombotic endotracheal occlusion in a mechanically ventilated patient. Crit Care. 2011, 15(2): 407.

20.Mannucci PM. Hemostatic drugs. N Engl J Med. 1998, 339(4): 245-253.

21.Mannucci PM. Hemostatic drugs. N Engl J Med. 1998, 339(4): 245-253.